The Observatories of the Carnegie Institution for Science (Carnegie Observatories), Pasadena, CA
Version 1.3
Christoph C. Birk (birk AT carnegiescience DOT edu)
Location of this document: http://instrumentation.obs.carnegiescience.edu/Software/ADC/adc.html
Please select the appropriate IP-numbers for the Acromags and
the name of the telescope operator computer; then press OK
to startup the main window.
If Auto Startup is selected on the
Preferences window, the Configuration
window will not be displayed and the GUI will use the settings
from the previous session.
Press the Run/Park button to run or park the ADC.
If the ADC is inserted but the Update button is not checked or the Control is off, there will be a warning icon at the center of the main window:
When you open the calibration window, you first select the Prism.
When you click “Start Calibration” the GUI
will start reading the voltages from the encoder and open window
with a plot of the raw readout and the current calculated
angle. The readout should be constant until you click the “+” or
“-“ button to start the motor.
Once the motor is running the values should change in a co/sinus
wave. Let it run until you have done at least two full revolutions;
the GUI will display the minimum and maximum values (in counts)
while the motor rotates and update a second plot window
with the current calculated angle (assuming the current min/max values).
Once the min/max values do not change anymore you should click the
“Stop calibration” button and the GUI will open a window with the
measured min/max values for the co/sinus curves.
That window will ask “Save Calibration?” .
Upto that point you may use the “Calibration” window as a debug-tool,
by just answering “No”, after “Stop Calibration”.
Once you have a valid calibration and press “Yes”, the GUI will store
the min/max values and use them for the angle calculation.
Now you need to adjust the ‘Phase’ so that the angle shows ‘0’ when the
dot is at the fiducial position (see above).
The preferences and the calibration are stored at /Users/Shared/Library/Preferences/edu.carnegiescience.obs.ADC
Once the "Control" has been turned on, the blue dot should be up
and the red dot pointing down when the telescope is at the zenith.
When the telescope is at 30 deg elevation (and below) both dots
should point in the same direction, perpendicular to the telescope tube.
Table courtesy of Dave Osip:
Elevation | A-prism | B prism |
90 | -180.0 | 0.0 |
80 | -173.9 | -6.1 |
60 | -159.7 | -20.3 |
50 | -149.7 | -30.3 |
40 | -134.3 | -45.7 |
30 | -90.0 | -90.0 |
2017-11-30, Christoph C. Birk, Carnegie Observatories