There are two Lantronix terminal servers (TCP/IP <-> serial converters): How-To set the IP number: 1) press reset-button for 5 seconds during power-on 2) arp -s xxx.xx.xxx.94 00:80:a3:23:9b:03 3) ping xxx.xx.xxx.94 (may take a while to respond) 4) telnet xxx.xx.xxx.94 (login as privileged user) define server ipaddress xxx.xx.xxx.94 5) power-cycle unit ####################################################################### 1) right-hand side IMACS electronics rack (imacsets16p) Port# Settings Device Notes 1 - special port unused 2 9600,8/1,N,XON f/4 DC-motors currently disabled 3 9600,8/1,N,XON - was Mosaic_1 flexure 4 38400,8/1,N,none MC-PC 5 38400,8/1,N,none HP-sensors 6 19200,8/1,N,none - was GP-375 7 19200,8/1,N,none - was GP-375 8 9600,8/1,N,none MMTF/CS100 etalon 9 9600,8/1,N,none GISMO/Linmot mask slide ######################################################################## 2) inside the black ABS-box inside IMACS (imacsets16r) Port# Settings Device Notes 1 - special port unused 2 19200,8/1,N,none f/4 LVDT focus-encoder Mosaic/3 3 19200,8/1,N,none f/2 LVDT focus-encoder Mosaic/2 4 9600,8/1,N,XON f/4 flexure E-515 piezo-controller (old) 5 9600,8/1,N,CTS f/2 flexure E-516 piezo-controller (new) 6 9600,8/1,N,none f/4 Omega Ionpump Mosaic/3 7 9600,8/1,N,none f/2 Omega Ionpump Mosaic/2 8 19200,8/1,N,none f/4 vacuum gauge GP-375 9 19200,8/1,N,none f/2 vacuum gauge GP-375 ####################################################################### xterm> telnet imacsets16r ... OR ... xterm> telnet imacsets16p Username> whatever Local_18> show port 1 Local_18> set priv Password> xxxxxx Local_18>> set port 1 flow none set port 1 speed 19200 set port 1 name save port 1 ... OR ... Local_18>> define port 1 flow none define port 1 speed 19200 define port 1 name Local_18>> log reboot device