IFUM Software Documentation - Release Notes
Version History
1.3,
1.2,
1.1,
1.0
v1.3 (2024-06-05)
- Added a "Calibrate" button to the 'IFU' pane on the IFUM hardhat window.
v1.2 (2024-05-08)
- Use the temperatures of the echelle and lores gratings for
the calculation of the auto-focus correction.
v1.1 (2024-03-27)
- Store the filter names in
Users/Shared/Library/Application Support/M2FS/filters.xml.
- The same file is used by M2FS.
- Updated the "Swap" icon to remind users to calibrate after swapping
the LoRes grism.
v1.0.1 (2023-11-15)
- Update the configuration label on the CCD window (and FITS record)
when saving a configuration file.
v1.0 (2023-06-16)
- Installed v1.0 on guanaco: /Applications/LCO/IFUM.app
- v0074 (2023-06-15)
- Updated the IFU icons.
- Added the occultor encoder positions to the setup files.
- Please remove the folder (below) to install the new defaults:
- ~/Library/Application Support/IFUM/Occultor
- v0073 (2023-06-08)
- Updated the IFU and lampsOff icons.
- Added an occultor icon for each of the 21 IFU/occultor combinations.
- Observing runs in February, March and May 2023
- v0071 (2022-10-03)
- Variable warning time when taking calibration lamp exposures
(3 times the Lamps-AutoOff time, defined on the
Preferences window).
- Read the DCU status (FF-Screen, calibration lamps) from the DCU-GUI
and write it to the FITS-header.
DCU-FFS = 'deployed' / DCU FF-Screen
DCU-LMPS= 'He Ar' / DCU lamps
- v0068 (2022-09-26)
- Read the ADC status (ADC-stage, ADC-Control, Mirror-Stage) from the ADC-GUI
and write it to the FITS-header.
ADC-STGE= 'in' / ADC stage
ADC-CTRL= 'auto' / ADC control
MIRROR = 'out' / Mirror stage
- v0067 (2022-09-26)
- Fixed a bug that did update "lastfile" even when the exposure was aborted.
- Added "1666.667 units/mm" to the IFU pane on the hardhat window.
- Ensure that there is a delay of at least 1 second between "SLIT" commands.
- v0066 (2022-09-23)
- Indicate the current slit position in the drop-down menu when in an invalid position.
- Eg. When IFU=STD, but the slit is in L80 the label will show "inv(L80)".
- v0065 (2022-09-23)
- Do no request the CRADLESTATE from the Beagle. Just assume both shoes are installed.
- v0064 (2022-09-23)
- Increase the maxiumum value for the calibration lamps to 100.
- Engineering run in Sep/2022
- v0063 (2022-06-09)
- Added 2 engineering values to the comment field of the
SLITNAME FITS record.
- Renamed the slits: { 300, 180, 80 }
- v0061 (2022-05-10)
- Increased the acceptable error in the occultor position from 5 to 25 units
- Set the default "nudge" on the occultor hardhat to 1000.
- v0059 (2022-05-09)
- Reverted to the M2FS LED-order: UV, Blue, Visible, N-Red, F-Red, IR
- Close GUI if not in IFUM mode.
- Comissioning run in May/2022
- v0058 (2022-05-02)
- Added the 'IFU' hardhat pane.
- v0052 (2022-03-16)
- Added the 'Shoes' hardhat pane.
- v0047 (2022-03-02)
- Fixed the transparency of the CCD-windows.
- v0046 (2021-12-03)
- Some bug-fixes.
- v0045 (2021-11-30)
- Re-order LEDs (
MCLED
command).
- v0042 (2021-09-29)
- Fixed the 'window snap' bug.
- v0038 (2021-09-23)
- Window snap fails on Jeb's and Mario's MacBook Pro.
- v0029 (2020-02-20)
- Testing with Jeb
- v0001 (2019-06-21)
- fork
M2FS
code
2023-06-16,
Christoph C. Birk, Carnegie Observatories