setpoint = 24.640 P=20.0, I=0.7 [1/h], D=5.0 [h], F=0.130+0.0360*(SetP-Tout)+"GTC" GTC = (glycol_supply-glycol_return)/2 - 0.25 The 'D' term appears larger than before because I removed an internal constant.At May/10 15:30 [CDT]: Revert to the "old" board.
At May/10 11:30 [CDT]: Pato replaces the ADC board for table sensor #2 ... same noise level.
Since May/8 18:00 [CDT]: T_Table = (T1+T2+T3)/3
Pato refills the glycol ... after about 8 hrs the table temperatures
start rising.
At May/7 13:30 [CDT]: PID: off
At May/7 11:20 [CDT]: PID: on
At May/7 3:15 [CDT]: PID: off
At May/6 12:00 [CDT]: I don't know what's going on ... glycol level is low
Since Apr/30 17:45 [CDT]:
P=20.0, I=0.7 [1/h], D=2.0 [h], F=0.130+0.0360*(SetP-Tout)-"GTC" GTC = 0.25 - (glycol_supply-glycol_return)/2