July/23:
Glycol: I=1.0
July/23:
The pump motor was replaced.
This reduced the frequency of "flushing" events but did not eliminate them.
The glycol supply remains about 0.15 [C] higher than the control value.
Since July/20:
The PLC temperature controller is acting up. The pump intermittently
switches to full power.
Under investigation ...
July/18 (17:30):
Re-enabled the meta-controller
July/18 (13:00):
Shut down the meta-controller on pfs@guanaco, testing for interference
from a "guanaco clone" ... divergence ... ie. there is no other meta-controller
running.
Since July/11:
Glycol: Pg=2.0, I=0.3, F=-0.400+0.0470*(SetP-Tout)+"GTC"Since June/29: